META TOPICPARENT name="AmsTracker"

Procedures for Tracker Power ON/OFF

Revision 222010-08-17 - AlbertoOliva

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for Tracker Power ON/OFF

Power On Procedure:

Changed:
<
<
  • NO Tracker switch on if TTCS is NOT running : Please follow instructions on how to operate the TTCS and how to monitor the Tracker temperatures before switching on the Tracker.
  • In the current DAQ Power Up procedure it will be asked you to switch on and make the TTCS working after the POWER STEP 2. The POWER STEP 3 corresponds to the Tracker switch on, please make this step only after that the TTCS is up and running.
  • Let a window open on AMI to check the Pump-Accumulator (Pump-SetPoint or Pt_01 Pt_02 in the AMI plots) temperature difference (has to be greater than 3 degrees) and to check the Tracker temperatures (< 30 degrees)
  • Power Step 3 aka Traker Switch ON: this task is made from the RunControl.
  • Data taking: calibration and taka taking runs are performed by the RunControl Procedures (WHISLIST, ...)
>
>
  • General Rule: NO TRACKER ON IF TTCS IS NOT RUNNING
  • In the current DAQ Power Up procedure it will be asked you to switch on and make the TTCS working after the Power Step 2. Please follow instructions on how to operate the TTCS and how to monitor the Tracker temperatures before switching on the Tracker. Let a window open on AMI to check the Pump-Accumulator (Pump-SetPoint or Pt_01 Pt_02 in the AMI plots) temperature difference (has to be greater than 3 degrees) and to check the Tracker temperatures (< 30 degrees.
  • The Power Step 3 corresponds to the Tracker switch on, please confirm the execution of this step to the Run Control only after that the TTCS is up and running.
  • Data taking: calibration and taka taking runs are performed by the RunControl Procedures (WHISLIST, ....
 
Deleted:
<
<

Update Tracker LogBook:

 

Power Off:

  • The Run Control Power Step 7 makes the whole job.
Added:
>
>

Update Tracker LogBook:

Revision 212010-08-16 - AlbertoOliva

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for Tracker Power ON/OFF

Power On Procedure:

Changed:
<
<
>
>
 
  • In the current DAQ Power Up procedure it will be asked you to switch on and make the TTCS working after the POWER STEP 2. The POWER STEP 3 corresponds to the Tracker switch on, please make this step only after that the TTCS is up and running.
Changed:
<
<
  • Let a window open on AMI to check the Pump-Accumulator (Pump-SetPoint or Pt_01 Pt_02 in the AMI plots) temperature difference (has to be greater than 3 degrees) and to check the Tracker temperatures (< 30 degrees)
>
>
  • Let a window open on AMI to check the Pump-Accumulator (Pump-SetPoint or Pt_01 Pt_02 in the AMI plots) temperature difference (has to be greater than 3 degrees) and to check the Tracker temperatures (< 30 degrees)
 
  • Power Step 3 aka Traker Switch ON: this task is made from the RunControl.
  • Data taking: calibration and taka taking runs are performed by the RunControl Procedures (WHISLIST, ...)

Update Tracker LogBook:

Added:
>
>

Power Off:

  • The Run Control Power Step 7 makes the whole job.

Revision 202010-08-16 - MatteoDuranti

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for Tracker Power ON/OFF

Line: 6 to 6
 
  • NO Tracker switch on if TTCS is NOT running: Please follow instructions on how to operate the TTCS and how to monitor the Tracker temperatures before switching on the Tracker.
  • In the current DAQ Power Up procedure it will be asked you to switch on and make the TTCS working after the POWER STEP 2. The POWER STEP 3 corresponds to the Tracker switch on, please make this step only after that the TTCS is up and running.
Changed:
<
<
  • Let a window open on AMI to check the Pump-Accumulator temperature difference (has to be greater than 3 degrees) and to check the Tracker temperatures (< 30 degrees)
>
>
  • Let a window open on AMI to check the Pump-Accumulator (Pump-SetPoint or Pt_01 Pt_02 in the AMI plots) temperature difference (has to be greater than 3 degrees) and to check the Tracker temperatures (< 30 degrees)
 
  • Power Step 3 aka Traker Switch ON: this task is made from the RunControl.
  • Data taking: calibration and taka taking runs are performed by the RunControl Procedures (WHISLIST, ...)

Update Tracker LogBook:

Deleted:
<
<
META FILEATTACHMENT attachment="configfile_4.dat" attr="" comment="JINF Configuration file (For T4)" date="1261604820" name="configfile_4.dat" path="JINFconfigfile_4.dat" size="185" stream="JINFconfigfile_4.dat" tmpFilename="/usr/tmp/CGItemp38260" user="mduranti" version="1"

Revision 192010-08-16 - AlbertoOliva

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for Tracker Power ON/OFF

Line: 13 to 13
 

Update Tracker LogBook:

Deleted:
<
<

Second Step Procedure:

  • From the directory /home/ams/JMDCController and after a standard calibration you can run
 
META FILEATTACHMENT attachment="configfile_4.dat" attr="" comment="JINF Configuration file (For T4)" date="1261604820" name="configfile_4.dat" path="JINFconfigfile_4.dat" size="185" stream="JINFconfigfile_4.dat" tmpFilename="/usr/tmp/CGItemp38260" user="mduranti" version="1"

Revision 182010-08-16 - AlbertoOliva

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"
Changed:
<
<

Procedures for tracker Normal Operations

>
>

Procedures for Tracker Power ON/OFF

 
Deleted:
<
<

The Philippe DSP page

The DSP Code Source

The Current TDR software is:

The Current JINF software is:

 

Power On Procedure:

Deleted:
<
<
  • CHECK TEMPERATURE STATUS: NO TRACKER SWITCH ON IF TTCS SYSTEM IS NOT RUNNING
    Operations to be performed:
  1. Mini-TTCS should be working on the secondary loop:
    • Main parameters to be checked (on Mini-TTCS control box, near Tracker PC in POC):
      • Gas Flow (blue panel)
      • Set Point T:(address 11)
      • Condenser T (address 13): should be at least 5 below Set Point

        Typical values: as Flow around 3 gr/s , Set Point : 24C , Condenser = 15C
  2. From PDS-5 Controller window on PCPOC01:
    • Switch-on TTPD-A to read temperature from the primary loop control electronics.
      (TTPD-B will not be used until the Mini-TTCS is installed)
  3. Load Q-list on JMDC:
    • Depending on which TTCE is powered ON give the command:
      • qlist_ctl start A
      • qlist_ctl start B
(... THIS STEPS ARE IN THE MAIN STARTUP PROCEDURE!!!)
  1. Verify the Q-list has been correctly loaded:
    • From “JMDC-A Controller” window on PCPOC02 : define node 013 (top corner left)
      and “get QQ-LIST”. (to open a new JMDC-A Controller window if not available, open a terminal window on PCPOC02 and type JMDC-A eas:hrdl pcgsc00)
  2. Check Tracker Temperatures:
  • TRACKER SWITCH-ON : ALL TASKS ARE UNDER RUN-CONTROL procedure
  1. Switch-on the 8 TPD on PDS-5 Controller window on PCPOC01
  2. Start-up procedure of JINF ( boot, load DSPCODE “29F2” through CMD file in JMDC)
  3. Start-up procedure of TDRS ( boot, load DSPCODE “29F3” through CMD file in JMDC)

Init Procedure:

  • TRACKER INITIALIZATION (i.e. BOOTING + CONFIGURATION):
  • Execution of the T_BOO.cmd from the JMDC-A controller (the initialization is not Basili compliand yet)
    • Daq-Initialization of JINF (set trigger to hold, set peculiar Power Group voltages, ...)
    • Daq-Initialization of TDR (load the configuration file 6001 for TDRs)

Calibration Procedure:

  • TRACKER CALIBRATION : ALL TASKS ARE UNDER RUN-CONTROL procedure (through WHISHLIST):
    • Be sure that GET_RESULT is performed by RUN-CONTROL in the wishlist controller after calibration otherways calibration data will not be recorded.
 
Changed:
<
<

DAQ Procedure:

  • TRACKER (AND ALL SUB-DETECTORS) DATA TAKING: ALL TASKS ARE UNDER RUN-CONTROL procedure (through WHISHLIST)
>
>
  • NO Tracker switch on if TTCS is NOT running: Please follow instructions on how to operate the TTCS and how to monitor the Tracker temperatures before switching on the Tracker.
  • In the current DAQ Power Up procedure it will be asked you to switch on and make the TTCS working after the POWER STEP 2. The POWER STEP 3 corresponds to the Tracker switch on, please make this step only after that the TTCS is up and running.
  • Let a window open on AMI to check the Pump-Accumulator temperature difference (has to be greater than 3 degrees) and to check the Tracker temperatures (< 30 degrees)
  • Power Step 3 aka Traker Switch ON: this task is made from the RunControl.
  • Data taking: calibration and taka taking runs are performed by the RunControl Procedures (WHISLIST, ...)
 

Update Tracker LogBook:

Changed:
<
<
>
>
 

Second Step Procedure:

  • From the directory /home/ams/JMDCController and after a standard calibration you can run
Line: 81 to 21
 
Deleted:
<
<

PaoloZuccon - 03-Dec-2009 *
 
META FILEATTACHMENT attachment="configfile_4.dat" attr="" comment="JINF Configuration file (For T4)" date="1261604820" name="configfile_4.dat" path="JINFconfigfile_4.dat" size="185" stream="JINFconfigfile_4.dat" tmpFilename="/usr/tmp/CGItemp38260" user="mduranti" version="1"

Revision 172010-01-13 - MatteoDuranti

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for tracker Normal Operations

Line: 43 to 43
 
  1. From PDS-5 Controller window on PCPOC01:
    • Switch-on TTPD-A to read temperature from the primary loop control electronics.
      (TTPD-B will not be used until the Mini-TTCS is installed)
  2. Load Q-list on JMDC:
Deleted:
<
<
 
    • Depending on which TTCE is powered ON give the command:
Changed:
<
<
      • ./TTCEAqlist_init
      • ./TTCEBqlist_init
>
>
      • qlist_ctl start A
      • qlist_ctl start B
 (... THIS STEPS ARE IN THE MAIN STARTUP PROCEDURE!!!)
  1. Verify the Q-list has been correctly loaded:
    • From “JMDC-A Controller” window on PCPOC02 : define node 013 (top corner left)
      and “get QQ-LIST”. (to open a new JMDC-A Controller window if not available, open a terminal window on PCPOC02 and type JMDC-A eas:hrdl pcgsc00)

Revision 162010-01-13 - MatteoDuranti

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for tracker Normal Operations

Line: 32 to 32
 
Deleted:
<
<
  • Notes:The DSP DAQ monitor name on flash is NOT compliant with "official" one, 29F2 (waiting to choose the right JINF program and to make the configuration files)
 

Power On Procedure:

  • CHECK TEMPERATURE STATUS: NO TRACKER SWITCH ON IF TTCS SYSTEM IS NOT RUNNING
    Operations to be performed:
Line: 58 to 57
 
  • TRACKER SWITCH-ON : ALL TASKS ARE UNDER RUN-CONTROL procedure
  1. Switch-on the 8 TPD on PDS-5 Controller window on PCPOC01
Changed:
<
<
  1. Start-up procedure of JINF ( boot, load DSPCODE “2932” through CMD file in JMDC)
>
>
  1. Start-up procedure of JINF ( boot, load DSPCODE “29F2” through CMD file in JMDC)
 
  1. Start-up procedure of TDRS ( boot, load DSPCODE “29F3” through CMD file in JMDC)

Init Procedure:

Revision 152010-01-13 - PaoloZuccon

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for tracker Normal Operations

Line: 8 to 8
 

The Current TDR software is:

  • Flash File name: 29F3
Changed:
<
<
  • Philippe File name: 3ef3 (changed on all TDRs the 17th Dec 2009)
>
>
  • Philippe File name: 3f13 a107_20100107
 
Line: 18 to 18
 
  • Notes: The DSP DAQ monitor name on flash is compliant with "official" one

The Current JINF software is:

Changed:
<
<
  • Flash File name: 2932
  • Philippe File name: 3932
  • Version date and number: 980A_20091112
>
>
  • Flash File name: 29F2
  • Philippe File name: 3992
  • Version date and number: A01D_201010113
 

Revision 142009-12-23 - MatteoDuranti

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for tracker Normal Operations

Line: 22 to 22
 
  • Philippe File name: 3932
  • Version date and number: 980A_20091112
  • Config File flash name: 7001
Changed:
<
<
>
>
 
  • Notes:The DSP DAQ monitor name on flash is NOT compliant with "official" one, 29F2 (waiting to choose the right JINF program and to make the configuration files)

Power On Procedure:

  • CHECK TEMPERATURE STATUS: NO TRACKER SWITCH ON IF TTCS SYSTEM IS NOT RUNNING
    Operations to be performed:
Line: 79 to 87
 
PaoloZuccon - 03-Dec-2009 *
Changed:
<
<
META FILEATTACHMENT attachment="paramfil.txt" attr="" comment="JINF Configuration file (generic)" date="1260547964" name="paramfil.txt" path="paramfil.txt" size="1380" stream="paramfil.txt" tmpFilename="/usr/tmp/CGItemp55594" user="pzuccon" version="2"
META FILEATTACHMENT attachment="configfile_4.dat" attr="" comment="JINF Configuration file (For T4)" date="1260547831" name="configfile_4.dat" path="configfile_4.dat" size="185" stream="configfile_4.dat" tmpFilename="/usr/tmp/CGItemp55520" user="pzuccon" version="1"
META FILEATTACHMENT attachment="paramfil_4.txt" attr="" comment="JINF Configuration file (For T4)" date="1260547977" name="paramfil_4.txt" path="paramfil_4.txt" size="1719" stream="paramfil_4.txt" tmpFilename="/usr/tmp/CGItemp55628" user="pzuccon" version="1"
>
>
META FILEATTACHMENT attachment="configfile_4.dat" attr="" comment="JINF Configuration file (For T4)" date="1261604820" name="configfile_4.dat" path="JINFconfigfile_4.dat" size="185" stream="JINFconfigfile_4.dat" tmpFilename="/usr/tmp/CGItemp38260" user="mduranti" version="1"

Revision 132009-12-23 - MatteoDuranti

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for tracker Normal Operations

Line: 10 to 10
 
  • Flash File name: 29F3
  • Philippe File name: 3ef3 (changed on all TDRs the 17th Dec 2009)
  • Config File flash name: 6001
Changed:
<
<
>
>
 
  • Notes: The DSP DAQ monitor name on flash is compliant with "official" one

The Current JINF software is:

  • Flash File name: 2932

Revision 122009-12-21 - AlbertoOliva

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for tracker Normal Operations

Line: 6 to 7
 The DSP Code Source

The Current TDR software is:

Changed:
<
<
  • Flash File name: 29F3

  • Philippe File name: 3ee3

  • Version date and number: 9C01_R04/20091201

>
>
  • Flash File name: 29F3
  • Philippe File name: 3ef3 (changed on all TDRs the 17th Dec 2009)
 
  • Config File flash name: 6001
Changed:
<
<
>
>
 
  • Notes: The DSP DAQ monitor name on flash is compliant with "official" one

The Current JINF software is:

Changed:
<
<
  • Flash File name: 2932

  • Philippe File name: 3932

  • Version date and number: 980A_20091112

>
>
  • Flash File name: 2932
  • Philippe File name: 3932
  • Version date and number: 980A_20091112
 
Line: 31 to 30
 
      • Condenser T (address 13): should be at least 5 below Set Point

        Typical values: as Flow around 3 gr/s , Set Point : 24C , Condenser = 15C
  1. From PDS-5 Controller window on PCPOC01:
    • Switch-on TTPD-A to read temperature from the primary loop control electronics.
      (TTPD-B will not be used until the Mini-TTCS is installed)
Changed:
<
<
  1. Load Q-list on JMDC:
    • login on pcpoc02 (ssh ams@pcpoc02)
    • cd ~/config
    • Depending on which TTCE is powered ON give the command:
      • ./TTCEAqlist_init
      • ./TTCEBqlist_init
>
>
  1. Load Q-list on JMDC:
    • login on pcpoc02 (ssh ams@pcpoc02)
    • cd ~/config
    • Depending on which TTCE is powered ON give the command:
      • ./TTCEAqlist_init
      • ./TTCEBqlist_init
(... THIS STEPS ARE IN THE MAIN STARTUP PROCEDURE!!!)
 
  1. Verify the Q-list has been correctly loaded:
    • From “JMDC-A Controller” window on PCPOC02 : define node 013 (top corner left)
      and “get QQ-LIST”. (to open a new JMDC-A Controller window if not available, open a terminal window on PCPOC02 and type JMDC-A eas:hrdl pcgsc00)
  2. Check Tracker Temperatures:
Changed:
<
<
>
>
 
  • TRACKER SWITCH-ON : ALL TASKS ARE UNDER RUN-CONTROL procedure
Line: 49 to 49
 
  1. Start-up procedure of TDRS ( boot, load DSPCODE “29F3” through CMD file in JMDC)

Init Procedure:

Changed:
<
<
  • TRACKER INITIALIZATION : ALL TASKS ARE UNDER RUN-CONTROL procedure (through XXXController and through our Exec_CF)
>
>
  • TRACKER INITIALIZATION (i.e. BOOTING + CONFIGURATION):
 
Changed:
<
<
  1. Daq-Initialization of JINF (set trigger to hold, set peculiar Power Group voltages, ...)
    ( = cd /home/ams/TrackerUser/JMDCCommander ./Exec_CF Initialization.txt, not Basili compliand yet)
  2. Daq-Initialization of TDR (load the configuration file 6001 for TDRs)
>
>
  • Execution of the T_BOO.cmd from the JMDC-A controller (the initialization is not Basili compliand yet)
    • Daq-Initialization of JINF (set trigger to hold, set peculiar Power Group voltages, ...)
    • Daq-Initialization of TDR (load the configuration file 6001 for TDRs)
 

Calibration Procedure:

  • TRACKER CALIBRATION : ALL TASKS ARE UNDER RUN-CONTROL procedure (through WHISHLIST):
Line: 61 to 62
 

DAQ Procedure:

  • TRACKER (AND ALL SUB-DETECTORS) DATA TAKING: ALL TASKS ARE UNDER RUN-CONTROL procedure (through WHISHLIST)

Update Tracker LogBook:

Changed:
<
<
  • Tracker short version of LogBook is in pcpoc08 under /home/ams/TrackerUser and his name is LogBook_From25Nov09.txt
    You can update this using this page: http://pcamss0.cern.ch/cgi-bin/mon/rc.o.cgi?queryDB04=Form
    and using "Find DataFile(s) : (eg 1073741826 or From-To)" form to look at the last data de-framed by Choutko software and removing the fields of no interest.
  • After updating the txt file on pcpoc08, please update also LogBook
>
>
 

Second Step Procedure:

Added:
>
>
  • From the directory /home/ams/JMDCController and after a standard calibration you can run
 
Changed:
<
<
-- PaoloZuccon - 03-Dec-2009
>
>

PaoloZuccon - 03-Dec-2009
  *

META FILEATTACHMENT attachment="paramfil.txt" attr="" comment="JINF Configuration file (generic)" date="1260547964" name="paramfil.txt" path="paramfil.txt" size="1380" stream="paramfil.txt" tmpFilename="/usr/tmp/CGItemp55594" user="pzuccon" version="2"

Revision 112009-12-13 - PaoloZuccon

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for tracker Normal Operations

Added:
>
>

The Philippe DSP page

The DSP Code Source
 

The Current TDR software is:

  • Flash File name: 29F3

  • Philippe File name: 3ee3

Revision 102009-12-11 - PaoloZuccon

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for tracker Normal Operations

The Current TDR software is:

Line: 65 to 65
  *

META FILEATTACHMENT attachment="paramfil.txt" attr="" comment="JINF Configuration file (generic)" date="1260547964" name="paramfil.txt" path="paramfil.txt" size="1380" stream="paramfil.txt" tmpFilename="/usr/tmp/CGItemp55594" user="pzuccon" version="2"
Deleted:
<
<
META FILEATTACHMENT attachment="configfile_0.dat" attr="" comment="JINF Configuration file (generic)" date="1260547801" name="configfile_0.dat" path="configfile_0.dat" size="165" stream="configfile_0.dat" tmpFilename="/usr/tmp/CGItemp55497" user="pzuccon" version="1"
 
META FILEATTACHMENT attachment="configfile_4.dat" attr="" comment="JINF Configuration file (For T4)" date="1260547831" name="configfile_4.dat" path="configfile_4.dat" size="185" stream="configfile_4.dat" tmpFilename="/usr/tmp/CGItemp55520" user="pzuccon" version="1"
META FILEATTACHMENT attachment="paramfil_4.txt" attr="" comment="JINF Configuration file (For T4)" date="1260547977" name="paramfil_4.txt" path="paramfil_4.txt" size="1719" stream="paramfil_4.txt" tmpFilename="/usr/tmp/CGItemp55628" user="pzuccon" version="1"

Revision 92009-12-11 - PaoloZuccon

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for tracker Normal Operations

The Current TDR software is:

Line: 6 to 6
 
  • Philippe File name: 3ee3

  • Version date and number: 9C01_R04/20091201

  • Config File flash name: 6001
Added:
>
>
 
  • Notes: The DSP DAQ monitor name on flash is compliant with "official" one

The Current JINF software is:

  • Flash File name: 2932

  • Philippe File name: 3932

  • Version date and number: 980A_20091112

Changed:
<
<
  • Config File flash name: not yet used
>
>
 
  • Notes:The DSP DAQ monitor name on flash is NOT compliant with "official" one, 29F2 (waiting to choose the right JINF program and to make the configuration files)

Power On Procedure:

  • CHECK TEMPERATURE STATUS: NO TRACKER SWITCH ON IF TTCS SYSTEM IS NOT RUNNING
    Operations to be performed:
Line: 58 to 62
 

Second Step Procedure:

-- PaoloZuccon - 03-Dec-2009 \ No newline at end of file

Added:
>
>
*

META FILEATTACHMENT attachment="paramfil.txt" attr="" comment="JINF Configuration file (generic)" date="1260547964" name="paramfil.txt" path="paramfil.txt" size="1380" stream="paramfil.txt" tmpFilename="/usr/tmp/CGItemp55594" user="pzuccon" version="2"
META FILEATTACHMENT attachment="configfile_0.dat" attr="" comment="JINF Configuration file (generic)" date="1260547801" name="configfile_0.dat" path="configfile_0.dat" size="165" stream="configfile_0.dat" tmpFilename="/usr/tmp/CGItemp55497" user="pzuccon" version="1"
META FILEATTACHMENT attachment="configfile_4.dat" attr="" comment="JINF Configuration file (For T4)" date="1260547831" name="configfile_4.dat" path="configfile_4.dat" size="185" stream="configfile_4.dat" tmpFilename="/usr/tmp/CGItemp55520" user="pzuccon" version="1"
META FILEATTACHMENT attachment="paramfil_4.txt" attr="" comment="JINF Configuration file (For T4)" date="1260547977" name="paramfil_4.txt" path="paramfil_4.txt" size="1719" stream="paramfil_4.txt" tmpFilename="/usr/tmp/CGItemp55628" user="pzuccon" version="1"

Revision 82009-12-11 - PaoloZuccon

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for tracker Normal Operations

The Current TDR software is:

Line: 24 to 24
 
  1. From PDS-5 Controller window on PCPOC01:
    • Switch-on TTPD-A to read temperature from the primary loop control electronics.
      (TTPD-B will not be used until the Mini-TTCS is installed)
  2. Load Q-list on JMDC:
Changed:
<
<
    • cd /home/ams/ttce
    • ./qlist_init
>
>
    • login on pcpoc02 (ssh ams@pcpoc02)
    • cd ~/config
    • Depending on which TTCE is powered ON give the command:
      • ./TTCEAqlist_init
      • ./TTCEBqlist_init
 
  1. Verify the Q-list has been correctly loaded:
    • From “JMDC-A Controller” window on PCPOC02 : define node 013 (top corner left)
      and “get QQ-LIST”. (to open a new JMDC-A Controller window if not available, open a terminal window on PCPOC02 and type JMDC-A eas:hrdl pcgsc00)
  2. Check Tracker Temperatures:

Revision 72009-12-10 - MatteoDuranti

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for tracker Normal Operations

The Current TDR software is:

Line: 49 to 49
 

DAQ Procedure:

  • TRACKER (AND ALL SUB-DETECTORS) DATA TAKING: ALL TASKS ARE UNDER RUN-CONTROL procedure (through WHISHLIST)
Added:
>
>

Update Tracker LogBook:

  • Tracker short version of LogBook is in pcpoc08 under /home/ams/TrackerUser and his name is LogBook_From25Nov09.txt
    You can update this using this page: http://pcamss0.cern.ch/cgi-bin/mon/rc.o.cgi?queryDB04=Form
    and using "Find DataFile(s) : (eg 1073741826 or From-To)" form to look at the last data de-framed by Choutko software and removing the fields of no interest.
  • After updating the txt file on pcpoc08, please update also LogBook
 

Second Step Procedure:

-- PaoloZuccon - 03-Dec-2009 \ No newline at end of file

Revision 62009-12-08 - MatteoDuranti

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for tracker Normal Operations

The Current TDR software is:

Line: 44 to 44
 
  1. Daq-Initialization of TDR (load the configuration file 6001 for TDRs)

Calibration Procedure:

Changed:
<
<
  • TRACKER CALIBRATION : ALL TASKS ARE UNDER RUN-CONTROL procedure (through WHISHLIST)
>
>
  • TRACKER CALIBRATION : ALL TASKS ARE UNDER RUN-CONTROL procedure (through WHISHLIST):
    • Be sure that GET_RESULT is performed by RUN-CONTROL in the wishlist controller after calibration otherways calibration data will not be recorded.
 

DAQ Procedure:

  • TRACKER (AND ALL SUB-DETECTORS) DATA TAKING: ALL TASKS ARE UNDER RUN-CONTROL procedure (through WHISHLIST)

Second Step Procedure:

Revision 52009-12-08 - MatteoDuranti

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for tracker Normal Operations

The Current TDR software is:

Line: 13 to 13
 
  • Version date and number: 980A_20091112

  • Config File flash name: not yet used
  • Notes:The DSP DAQ monitor name on flash is NOT compliant with "official" one, 29F2 (waiting to choose the right JINF program and to make the configuration files)
Changed:
<
<

Power On Procedure

Init Procedure

Calibration Procedure

Second Step Procedure

>
>

Power On Procedure:

  • CHECK TEMPERATURE STATUS: NO TRACKER SWITCH ON IF TTCS SYSTEM IS NOT RUNNING
    Operations to be performed:
  1. Mini-TTCS should be working on the secondary loop:
    • Main parameters to be checked (on Mini-TTCS control box, near Tracker PC in POC):
      • Gas Flow (blue panel)
      • Set Point T:(address 11)
      • Condenser T (address 13): should be at least 5 below Set Point

        Typical values: as Flow around 3 gr/s , Set Point : 24C , Condenser = 15C
  2. From PDS-5 Controller window on PCPOC01:
    • Switch-on TTPD-A to read temperature from the primary loop control electronics.
      (TTPD-B will not be used until the Mini-TTCS is installed)
  3. Load Q-list on JMDC:
    • cd /home/ams/ttce
    • ./qlist_init
  4. Verify the Q-list has been correctly loaded:
    • From “JMDC-A Controller” window on PCPOC02 : define node 013 (top corner left)
      and “get QQ-LIST”. (to open a new JMDC-A Controller window if not available, open a terminal window on PCPOC02 and type JMDC-A eas:hrdl pcgsc00)
  5. Check Tracker Temperatures:
  • TRACKER SWITCH-ON : ALL TASKS ARE UNDER RUN-CONTROL procedure
  1. Switch-on the 8 TPD on PDS-5 Controller window on PCPOC01
  2. Start-up procedure of JINF ( boot, load DSPCODE “2932” through CMD file in JMDC)
  3. Start-up procedure of TDRS ( boot, load DSPCODE “29F3” through CMD file in JMDC)

Init Procedure:

  • TRACKER INITIALIZATION : ALL TASKS ARE UNDER RUN-CONTROL procedure (through XXXController and through our Exec_CF)
  1. Daq-Initialization of JINF (set trigger to hold, set peculiar Power Group voltages, ...)
    ( = cd /home/ams/TrackerUser/JMDCCommander ./Exec_CF Initialization.txt, not Basili compliand yet)
  2. Daq-Initialization of TDR (load the configuration file 6001 for TDRs)

Calibration Procedure:

  • TRACKER CALIBRATION : ALL TASKS ARE UNDER RUN-CONTROL procedure (through WHISHLIST)

DAQ Procedure:

  • TRACKER (AND ALL SUB-DETECTORS) DATA TAKING: ALL TASKS ARE UNDER RUN-CONTROL procedure (through WHISHLIST)

Second Step Procedure:

  -- PaoloZuccon - 03-Dec-2009 \ No newline at end of file

Revision 42009-12-07 - MatteoDuranti

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for tracker Normal Operations

Changed:
<
<

The Current JINF software is:

>
>

The Current TDR software is:

 
  • Flash File name: 29F3

  • Philippe File name: 3ee3

  • Version date and number: 9C01_R04/20091201

  • Config File flash name: 6001
  • Notes: The DSP DAQ monitor name on flash is compliant with "official" one
Changed:
<
<

The Current TDR software is:

>
>

The Current JINF software is:

 
  • Flash File name: 2932

  • Philippe File name: 3932

  • Version date and number: 980A_20091112

Revision 32009-12-07 - MatteoDuranti

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for tracker Normal Operations

The Current JINF software is:

Changed:
<
<
  • Flash File name:

  • Philippe File name:

  • Version date and number:

  • Config File flash name:
  • Notes:
>
>
  • Flash File name: 29F3

  • Philippe File name: 3ee3

  • Version date and number: 9C01_R04/20091201

  • Config File flash name: 6001
  • Notes: The DSP DAQ monitor name on flash is compliant with "official" one
 

The Current TDR software is:

Changed:
<
<
  • Flash File name:

  • Philippe File name:

  • Version date and number:

  • Config File flash name:
  • Notes:
>
>
  • Flash File name: 2932

  • Philippe File name: 3932

  • Version date and number: 980A_20091112

  • Config File flash name: not yet used
  • Notes:The DSP DAQ monitor name on flash is NOT compliant with "official" one, 29F2 (waiting to choose the right JINF program and to make the configuration files)
 

Power On Procedure

Init Procedure

Calibration Procedure

Revision 22009-12-03 - PaoloZuccon

Line: 1 to 1
 
META TOPICPARENT name="AmsTracker"

Procedures for tracker Normal Operations

T he Current JINF software is:

Revision 12009-12-03 - PaoloZuccon

Line: 1 to 1
Added:
>
>
META TOPICPARENT name="AmsTracker"

Procedures for tracker Normal Operations

T he Current JINF software is:

  • Flash File name:

  • Philippe File name:

  • Version date and number:

  • Config File flash name:
  • Notes:

The Current TDR software is:

  • Flash File name:

  • Philippe File name:

  • Version date and number:

  • Config File flash name:
  • Notes:

Power On Procedure

Init Procedure

Calibration Procedure

Second Step Procedure

-- PaoloZuccon - 03-Dec-2009

 