Outer Tracker Scanning System in Situ

Before and after irradiations reference measurements are performed with a scanning system of two 74 MBq 90Sr sources transported by a step motor.

Scanning System in Situ

Main Elements:

  • Scanning Frame

scanning_system.JPG

  • 2 90Sr sources of 74MBq, moved by a step motor

sources.JPG

  • Current-meter and Safety Garage

current_meter.JPG

Schematic of the scanning source:

schematic.jpg

How to run the scan

For the Sr scans you need the portable computer OTSCAN01W.

The executable files can be found in: C:\Documents and Settings\Administrator\My Documents\CVI-programs. The file we need to start the scan of one module is clrscan.exe. We test the front layer and the back layer (the source is shining through the front layer). One layer is tested at a time.

As the front end boards do not measure current (but drift times), a custom made Ampere meter box is connected replacing temporarily the front end of the module to be tested. As this is a very delicate operation, only a limited number of people are allowed to do it.

Give source height : 4 for front layer, 4.5 for the back layer. Give module name: production/ serial number of the module. Both sides are A and B but there' s no rule which one is A and which B. Give high voltage:1600

Then click on "Accept to continue".

The system will prepare.

Check if all the settings are alright. If yes, click on "Accept to continue" in the second window. Then, click on "Start scan" in the new pop-up window. Even if the settings are not right, start the run, and change the settings manually from the CAEN menu. i.e. for high voltage setting push F 1 and the nominal voltage value.

The logbook is kept on a pluscc node in /calib/ot/Scanning tool/logbooks. All the scan logs are kept in one text file: logbook-scanning.txt.

The files have to be copied manually form OTSCAN01W (C:\Test\Srscan\$Module Number$) to a pluscc node (/calib/ot/ScanningTool/SrScan/$ModuleNumber$).

You can choose to monitor the current in a given wire, or altogether (type 65, then Enter in "Wire choice" field). SrScan.png

When someone is working with the FE boxes, make sure the source is as high as possible not to create unnecessary risk for the person working. You can use "usemotor.exe" , "Go to positive limit switch". When the source is high enough ~the beam pipe, "Stop motor at once".

Analysis

The analysis is typically carried out comparing the current behavior after and before an irradiation step. The ratio plot produced by this comparison is normalised using the average value obtained inside a region where the source is not present. A zone around the source, “source region”, is selected to study in details the effect of the irradiation. Two quantities are proposed to estimate the ageing effects: the average of the two absolute minima in the source region and the “region factor” quatified summing the deviation from 1 of all the cells inside the source region and dividing by the number of cells considered. If there are no gain changes in the irradiated area these two quantities must be 1 and 0 respectively.

How To Run it

  • Connect to the pluscc node with "ot_oper" user

  • Run the script: SrScan.sh [-t -n -f -m -s] <ModuleNumber>
    • -t list of irradiation hours (useful only if you enable "trend_plot" function in the option file)
    • -n installation setup ('cern' for first monolayer, 'cern2' for second monolayer)
    • -m detector part scanned ('L_half' for the lower part of the monolayer, 'U_half' for the upper part of the monolayer, 'all' for the entire detector)
    • -f force to set new input file for the analysis
    • -s run single scan analysis

  • Examples:
    • SrScan.sh -t 64 -f -s 90

  • Output save in:
    • /scratch/everyone/ot_oper/SrScan/<ModuleNumber>/output/plots
    • or in /calib/ot/ScanningTool/SrScan/<ModuleNumber>/output/plots

-- BarbaraStoraci - 12 Mar 2009

Topic attachments
I Attachment History Action Size Date WhoSorted ascending Comment
JPEGjpg current_meter.JPG r1 manage 51.3 K 2009-03-12 - 14:18 BarbaraStoraci Current-meter
JPEGjpg scanning_system.JPG r1 manage 152.3 K 2009-03-12 - 14:21 BarbaraStoraci Scanning System in Situ
JPEGjpg schematic.jpg r1 manage 15.9 K 2009-03-12 - 16:11 BarbaraStoraci Schematic of the source system
JPEGjpg sources.JPG r1 manage 39.4 K 2009-03-12 - 16:08 BarbaraStoraci Scanning Sources
PNGpng SrScan.png r1 manage 59.6 K 2012-01-10 - 16:36 EvelinaMarinova  
Edit | Attach | Watch | Print version | History: r6 < r5 < r4 < r3 < r2 | Backlinks | Raw View | WYSIWYG | More topic actions
Topic revision: r6 - 2017-05-19 - EvelinaMarinova
 
    • Cern Search Icon Cern Search
    • TWiki Search Icon TWiki Search
    • Google Search Icon Google Search

    LHCb All webs login

This site is powered by the TWiki collaboration platform Powered by PerlCopyright & 2008-2019 by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding TWiki? Send feedback