Check Calibration

A calibration is delivered at the beginning of each run (every 30 min.). Please check it using the program MonitorUI.

Screenshot-Monitor.png

  1. If MonitorUI is open press the Update button or set the auto-update box.
  2. If MonitorUI is not open launch it with the command:
    ssh tracker@pcpoc08 MonitorUI TrackerUser/Monitor/Output/LastCalDB.root
    
  3. Once you have a calibration compare it with the reference one (see below). Please check:
    • Overflows in the two leftmost plots should be exactly 0 (upper-blue) and 385 (lower-red).
    • The middle plots graph are peaked at approximately 2.7 (upper-blue) and 3.3 (lower-red) ADC counts.
    • The rightmost plots values are similar to the one seen in reference. In particular bits from 5 to 14 must be 0.
  4. If one of the previous is true is possible that the calibration failed. Ask to make another calibration.
  5. If the problem is still there ask to an expert.

Reference Calibration Summary:

Reference_Cal2.png

Topic attachments
I Attachment History Action Size Date Who Comment
PNGpng Reference_Cal.png r1 manage 40.9 K 2010-10-01 - 21:21 AlbertoOliva Reference Calibration
PNGpng Reference_Cal2.png r1 manage 46.6 K 2010-10-01 - 22:25 AlbertoOliva Reference Calibration with Comments
PNGpng Screenshot-Monitor.png r1 manage 11.4 K 2010-10-01 - 22:17 AlbertoOliva Screenshot of the MonitorUI with comment
Edit | Attach | Watch | Print version | History: r1 | Backlinks | Raw View | WYSIWYG | More topic actions
Topic revision: r1 - 2010-10-01 - AlbertoOliva
 
    • Cern Search Icon Cern Search
    • TWiki Search Icon TWiki Search
    • Google Search Icon Google Search

    Main All webs login

This site is powered by the TWiki collaboration platform Powered by PerlCopyright & 2008-2019 by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding TWiki? Send feedback