ProbeStation
CERN HGTD Probe Station
Introduction
This twiki contains instructions for using the HGTD probe station to make IV or CV measurements.
Location: 161/1-006
Equipment
Experimental setup
- A Keithley 2410 source meter (on loan from the CERN ITk Group, who have our thanks!)
- A Keithley 6487 picoammeter (on loan from the CERN Epool)
- A micro PC (Windows 10/Linux)
- A probe station (on loan from the CERN ITk Group)
- A vacuum pump (to ensure that the sensor remains in place on the chuck)
- XIMEA camera to take pictures of the sensor through the microscope
- You can install the camera software here (already installed on lab PC): https://www.ximea.com/support/wiki/allprod/XIMEA_CamTool
- To open the software click the xiCamTool shortcut on the desktop (applies in both Windows and Linux)
- Make sure the light is on (and the source meter is off!) in the probe station or no image will be visible!
- Measurements are controlled using a labview project (Windows only) developed for the probe station setup used by RD-50.
Pin controls
- Pins are used to supply the HV, ground the guard ring and connect the sensor pad, as shown in the left image below.
- The pins are connected as follows (as illustrated in the lower right schematic):
- the pin closest to the cabinet door is to ground the guard ring,
- the middle pin provides the HV to the chuck,
- the further pin is to measure the current through the sensor pad.

- To position the pins on the pad or guard ring is tricky. Look through the microscope and use the small knobs on the needle mounts.
- The function of each of these knobs is shown in the schematic below:
Instructions for measurements
The HGTDSensors twiki contains instructions/recommendations for making sensor tests.
Currently we're keeping the CERN team measurements in this spreadsheet
.
To set up the PC for IV or CV measurements:
- Switch on computer, booting up Windows
- Load up labview project:
- Documents → Labview project → cviv-daq → CVIV
- Select CV-IV-control.vi from the list of options displayed in LLB Manager
- When the CV-IV-control.vi project opens, click on the
,
button on the menu bar
- This takes you from edit mode to running mode
- Fill out the relevant fields in the labview GUI according to the measurements you wish to make (see following sections).
IV (floating guard ring)
Labview settings
Parameter |
Setting |
Comment |
Integration time |
3 [20 ms] |
Drop down menu to right of "Device name" |
pFZ |
Default assuming non-irradiated sensors. |
Number of steps |
Calculate (2V) |
e.g. Start: 0, Stop: -100, Number of steps: 50 |
Tester |
|
|
Start |
0 V |
|
Step mode |
linear |
|
Averaging (count) |
3 |
Number of measurements taken at a single voltage point for averaging. |
GPIB |
21 |
Port for Keithley 2410 |
Type of measurement |
IV |
Select from drop-down menu. |
Compliance. |
1e-6 A |
Sensor dependent (have seen nA currents for HPK Type 1) |
Compl. Stop |
ON |
This ensures that the measurement cycle will cease, and the voltage ramp down, should the sensor reach the compliance you set. |
Stop |
-100 V |
This is the max voltage for the scan (will differ between sensors). Note that the value must be negative. |
Device name |
Sensor name |
Usually this should be the name written on the sensor packaging. This will be the name of the folder in which the results will be stored. |
IV (grounded guard ring)
Connecting the sensor
- Ensuring the voltage source is off, open the probe station cabinet and switch on the light.
- The light switch is located in the back left-hand corner of the cabinet.
- Turn on the vacuum pump.
- Turbulence from the pump the instant it is switched on will cause the sensor to move violently if you try to turn on the pump after placing the sensor!
- Using plastic tweezers, gently place the sensor onto the chuck, positioning it over the small hole in the centre.
- The sensor should completely cover the hole so that the vacuum is effective.
- Position the needles on the sensor, as shown below.
- Switch off the light and close the cabinet doors.
Labview settings
Parameter |
Setting |
Comment |
Device name |
Sensor name |
Usually this should be the name written on the sensor packaging. This will be the name of the folder in which the results will be stored. |
Drop down menu to right of "Device name" |
pFZ |
Default assuming non-irradiated sensors. |
Tester |
|
|
Type of measurement |
IV with two I-meter |
Select from drop-down menu. |
Start |
0 V |
|
Stop |
-100 V |
This is the max voltage for the scan (will differ between sensors). Note that the value must be negative. |
Number of steps |
Calculate (2V) |
e.g. Start: 0, Stop: -100, Number of steps: 50 |
Step mode |
linear |
|
Compl. Stop |
ON |
This ensures that the measurement cycle will cease, and the voltage ramp down, should the sensor reach the compliance you set. |
V source and I meter |
Keithley 2410 |
Select from drop-down menu. |
I meter |
Keithley 6487 |
Select from drop-down menu. |
GPIB address (Keithley 24xx) |
21 |
Port for Keithley 2410 |
GPIB address (Keithley 6487) |
?? |
Port for Keithley 6487 |
Integration time |
3 [20 ms] |
Averaging (count) |
3 |
Number of measurements taken at a single voltage point for averaging. |
Compliance. |
1e-6 A |
Sensor dependent (have seen nA currents for HPK Type 1) |
Analysis of IV/CV data:
Useful links
-- EmmaKuwertz - 2019-05-21
This topic: Sandbox
> WebPreferences > ProbeStation
Topic revision: r2 - 2019-07-10 - EmmaKuwertz